FD-9 Auto Scan Spectrophotometer

  • Printer
Category: ProductsColor
Model FD-9
Illumination/viewing system 45°a: 0° (annular illumination)
Conforms to CIE No. 15, ISO 7724/1, DIN 5033 Teil 7, ASTM E 1164, and JIS Z 8722 Condition a for reflectance measurements
Spectral separation device Concave grating
Wavelength range 380 nm to 730 nm
Wavelength pitch 10 nm
Half bandwidth Approx. 10 nm
Measurement area Approx Φ3 mm
Light source LED
Measurement range Reflectance: 0 to 150%
Short-term repeatability Within σΔE00 0.05
(When white plate is measured 30 times at 10-second intervals after white calibration has been performed)
Inter-instrument agreement Within ΔE00 0.3 (Average of 12 BCRA Series II color tiles compared to values measured with a master body under Konica Minolta standard conditions)
Measurement time Approx 4 min./1,500 patch
(Under Konica Minolta measurement condition)
Output data Spectral reflectance data
Measurement conditions*2 Corresponding to ISO 13655 Measurement Conditions M0 (CIE Illuminant A), M1 (CIE Illuminant D50), and M2 (illumination with UV-cut filter); User-defined illuminant
Sample backing Corresponding to ISO 13655 Standard white backing
Interface USB2.0, 100Base-TX
Power Included AC adapter: Input: AC 100 to 240V 50-60Hz; Output: DC 24V 5A
Dimensions 716×251×159mm(Body only)
(W × D × H) 601×1158×245mm(Auto sheet feeder only)
Weight 10.5kg(Body only)
Operating temperature/ humidity range 10 to 35°C, 30 to 85% relative humidity with no condensation
Storage temperature/humidity range 0 to 45°C, 0 to 85% relative humidity with no condensation
Standard accessories AC adapter,USB cable,Measurement Utility software FD-S2w
Optional accessory Auto sheet feeder